Title: A comparative study of intraocular pressure measurements using Goldmann Applanation Tonometer, I-care pro and Corvis ST after DSAEK
Author(s): Farhad Nejat, Khosrow Jadidi
Presentation Type: Poster
Subject: Cornea & lens
Others:
Presenting Author:
Name: Khosrow Jadidi
Affiliation :(optional) Vision health Research Center, Semnan University of medical sciences
E mail: kh.jadidi@gmail.com
Phone:
Mobile: 09121053475
Purpose:

To compare the intraocular pressure (IOP) measurements using different devices and evaluate the influence of central corneal thickness (CCT) on IOP measurements following DSAEK surgery.

Methods:

In this cross-sectional study, 22 eyes from 19 patients (mean age: 56.59 years) after DSAEK with clear and compact graft were enrolled. Mean follow-up after the surgery was 2.191±1.852 years. IOP was measured with GAT, I-care pro and Corvis ST (as non-corrected and biomechanical IOP) in post-DSAEK patients and agreement between measurements of three devises were assessed. Corvis ST was used to measure the CCT value. The influence of CCT on IOP measurements was evaluated using spearman test.

Results:

The mean IOP measured by GAT, I-care pro and Corvis as the non-corrected and biomechanical IOP were 12.545±2.064, 11.591±2.465, 12.686±3.967 and 9.714±4.339 respectively. GAT measurement was considered as a gold standard and the only significant difference was observed between GAT and Corvis biomechanical IOP (P<0.05). CCT did not correlate with any of the IOP measurements except Corvis biomechanical IOP (r= -0.518, P=0.01).

Conclusion:

There was a good agreement between GAT, I care and Corvis non-corrected IOP measurements while the Corvis biomechanical IOP measurement was less than the GAT.

Attachment: 60IOP after DSAEK poster.pptx